Industry-recognized accuracy standard. The ±20% / ±20 mg/dL "20/20 rule" used below is a widely cited clinical accuracy benchmark for blood-glucose and CGM devices — the kind of threshold regulatory bodies including the FDA reference when evaluating and clearing home glucose-monitoring devices.

BG vs SG Accuracy Check

Quick 20/20 rule calculator for Technical Support
MiniMed Support Tool
BG (fingerstick)
0
SG (sensor)
0
⌨ TYPE NUMBERS · TAB = SWITCH · ENTER = CHECK · ⌫ = DELETE

What am I checking?

BG (Blood Glucose) is the fingerstick meter reading — a direct capillary blood measurement, treated as the reference value. SG (Sensor Glucose) is the CGM's reading, taken from interstitial fluid. The two normally differ a little — interstitial fluid lags blood by roughly 5–15 minutes — so some gap is expected, especially right after a meal, dosing, or exercise when glucose is changing quickly.

The 20/20 rule

A widely used clinical accuracy benchmark for glucose-monitoring devices: a device reading is considered accurate if it falls within ±20% of the reference value at higher glucose levels, or within ±20 mg/dL at lower glucose levels where a percentage would be too tight to be meaningful. Regulatory bodies including the FDA have used thresholds like this when evaluating and clearing home blood-glucose and CGM devices.

BG readingRule applied
Below 80 mg/dL (4.4 mmol/L)SG must be within ±20 mg/dL of BG
80 mg/dL and aboveSG must be within ±20% of BG

This calculator uses 80 mg/dL as the cutover point. Some references use 70 or 75 mg/dL instead — the underlying idea is the same: below that point, a fixed mg/dL window is used instead of a percentage.

How to use this with a PT call

Ask the patient for their current fingerstick (BG) and what the sensor is showing (SG) at the same moment, enter both here in the unit they're using, and tap Check accuracy. A PASS means the sensor is within the expected accuracy range for its current level — some gap is normal. A FAIL means the two readings are farther apart than the 20/20 rule allows, which may warrant a sensor check, recalibration, or escalation per your usual troubleshooting flow.

This is a quick reference check, not a diagnostic or clinical decision tool. It does not replace manufacturer calibration algorithms (which are proprietary) or your organization's official troubleshooting and escalation procedures. Brand colors reference Medtronic's published palette for internal presentation purposes only.

Why this exists: there wasn't an official quick-check tool like this available internally, so I built one — this is one of a handful of self-initiated support tools I've made to fill small gaps like this.